Displacement Transducers
3-Channel Transducer Indicator
Strain
- Full Bridge 350E and above.
Strain Gauge Transducers
- Strain gauge-based Transducers.
- Load Cells.
- Pressure Cells.
- Torque Transducers.
- LCD Display.
- Resolution (Factory set)
- 1 in 1000, 2000, 3000, 5000, or 10000 (as per
quote) - Tare Facility (Manual & Software).
Applications
- Force / Load Cell output Measurements.
- Pressure Measurement.
- Torque Measurement.
- Strain Measurement.
Features
- Uses “HIGH PERFORMANCE INSTRUMENTATION AMPLIFIER”.
- 5ppm, 0.1% Accuracy Resistors provide excellent calibration.
- Short circuit protected output.
Exceptional Linearity, stability and repeatability.
| TECHNICAL SPECIFICATIONS | |
|---|---|
| Number of channels for measuring input data | 3 Channels |
| Digital Display | Alpha-Numeric 16-digit X 2 lines LCD Dot Matrix Display |
| Applicable gauge resistance | Quarter Bridge:120 ohm & 350 ohm. Half & Full Bridge: 120-1000 ohm |
| Bridge excitation | DC 2V |
| Gauge Factor Setting | Yes, Through 10 Turn Pot with Lockable Dial |
| Measuring range | ±20000×10-6 strain (bridge excitation DC 2V) |
| Calibration | Built-in Shunt Cal facility |
| Balancing method (Each Channel) | Through Ten Turn Variable Pot |
| Balancing accuracy | within ±3×10-6 strain |
| Stability on zero | ±1×10-6 strain/°C ±1digit |
| Power supply | AC, 230V ±10% at 50Hz |
| Operating Temperature | 0 to 50 °C |